Up to top level
AO15   AO16   AO17   AO18   Backgrounds   Calibration   Conference   Data   Docs   EPICMOS   EPICpn   Feedback   Gallery   Misc   OM   Pending   PhD_Theses   Publications   RGS   RadMonitor   SAS_Hardware   SAS_WS   SASv16.0   SASv16.0_Installation   SASv16.1   SASv16.1_Installation   SASv17.0   SASv17.0_Installation   SASv18.0   SASv18.0_Installation   SciSim   Simulators_other   Suggestions   Trash   Visibility   XMM-bouncing   XMM-news   XRPS   XSA   esas   incoming  

Logged in as guest

Viewing SASv16.0_Installation/82016
Full headers

From: nadine.may7@gmail.com
Subject: EPICpn
Compose reply
Download message
Move To:
1 replies: 1
0 followups:

Private message: yes  no

Notes:

Notification:


Date: Thu, 20 Jul 2017 13:13:57 GMT
From: nadine.may7@gmail.com
To: xmmhelp@sciops.esa.int
CC: nadine.may7@gmail.com
Subject: EPICpn
Full_Name: Nadine
Submission from: (NULL) (137.158.93.48)


In running epproc, the command seems to fail.Below is the message it outputs.
Below the output in another output with higher verbosity.
I have ran the epproc command on 2 different installations of sas and the same
output is given in both.  

** epproc::hkgtigen: warning (Mismatch), EMOS1/CONTAMINATION is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (Mismatch), EMOS1/FWC is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (Mismatch), EMOS1/SCREENTHRESH is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (Mismatch), EMOS2/CONTAMINATION is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (Mismatch), EMOS2/FWC is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (Mismatch), EMOS2/SCREENTHRESH is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (Mismatch), EPN/BUCLOSEDODI is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (Mismatch), EPN/FWC is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (Mismatch), EPN/TICLOSEDODI is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (Mismatch), EPN/TIMEJUMPTOL is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (Mismatch), RGS1/SAACORR is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (Mismatch), RGS2/SAACORR is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (Mismatch), XMM/SPECQUAL is not a known
instrument/filetype combination
** epproc::hkgtigen: warning (EmptyGtiSet), The generated GTI data set
`.//1479_0510011601_EPN_HkGti.ds' is empty - applying it to any event list will
result in no events getting selected.

Higher verbosity output Below:

epproc:- Executing (routine): epproc removetemporaries=yes
removeintermediategtis=yes removeintermediateeventlists=yes selectccds=no
ccd1=no ccd2=no ccd3=no ccd4=no ccd5=no ccd6=no ccd7=no ccd8=no ccd9=no ccd10=no
ccd11=no ccd12=no selectmodes=yes imaging=yes timing=yes burst=no
timingsrcposition=190 withinstexpids=no instexpids='' withgtiset=no
gtiset=gti.ds runhkgtigen=yes runatthkgen=yes referencepointing=median ra=0
dec=0 posangle=0 filterevents=yes filterexpression='(PI > 150 && RAWY
> 0)'
flagfilteredevents=no rungtimerge=no applygti=yes runevlistcomb=yes
runepexposure=yes deleteexposurecolumns=yes searchforbadpixels=yes
searchforbadcolumns=yes thresholdlabel=rate lothresh=0 hithresh=0.0045
columnsearchlabel=median locolthresh=0 hicolthresh=0.00105 flickertimesteps=1
flickerksthresh=0.55 flickerchisqthresh=15 backgroundrate=0.0001
narrowerthanpsf=1.5 threshabovebackground=no loenergythresh=0.14
hienergythresh=12 randomizeposition=yes randomizeenergy=yes gainctiaccuracy=2
reemissionthresh=0 withoutoftime=no mappatterntype=sssd wrongpixlimit=20
setupbpx=nom3 lowerthreshold=20 withsrccoords=no srcra=0 srcdec=0
mipthreshold=3000 mipmethod=onboard mipdist=no mipdiscard=yes withparameters=no
parameters='' except=no withoverrideparameters=no overrideparameters=''
analyzingSciSimdata=no runepreject=no badcolumnset=badcolumns.tab sigma=4
noiseparameters='0.98 1 1 1 1 1 1 1 1 1 1 1 1' withoffsetmap=yes set=odfset
withsoftflarescreening=no softflarethreshold1=10 softflarethreshold2=1
softflaresmooth=BOX softflareenergyrange='40 50' softflaresmoothparams='2 1 1' 
-V 5
epproc:- epproc (epicproc-2.9.13)  [xmmsas_20060628_1801-7.0.0] started: 
2017-07-20T12:15:13.000
epproc:- Selected data mode: 0 [Imaging]
epproc:- Selected data mode: 1 [Timing]
epproc:- Selected CCD: 1
epproc:- Selected CCD: 2
epproc:- Selected CCD: 3
epproc:- Selected CCD: 4
epproc:- Selected CCD: 5
epproc:- Selected CCD: 6
epproc:- Selected CCD: 7
epproc:- Selected CCD: 8
epproc:- Selected CCD: 9
epproc:- Selected CCD: 10
epproc:- Selected CCD: 11
epproc:- Selected CCD: 12
epproc:- 0 [Imaging] -> 0
epproc:- 1 [Timing] -> 0
epproc:- Considering one of the 0 exposures for mode 0 [Imaging]
epproc:- Considering one of the 0 exposures for mode 1 [Timing]
epproc:- .//1479_0510011601_AttHk.ds exists. atthkgen will not be run again.
epproc:- Executing (invoked): hkgtigen instrument=epn
gtiset=.//1479_0510011601_EPN_HkGti.ds diagfile=STDOUT plotgtis=no
diagnosticmode=no withgtiset=yes parameters='' withparameters=no except=no
overrideparameters='' overrideparametervalues='' withoverrideparameters=no
timecolumn=TIME prefraction=0.5 postfraction=0.5  -V 5
epproc:- hkgtigen (hkgtigen-1.14.2)  [xmmsas_20060628_1801-7.0.0] started: 
2017-07-20T12:15:13.000
epproc::hkgtigen:- Executing (routine): hkgtigen instrument=epn
gtiset=.//1479_0510011601_EPN_HkGti.ds diagfile=STDOUT plotgtis=no
diagnosticmode=no withgtiset=yes parameters='' withparameters=no except=no
overrideparameters='' overrideparametervalues='' withoverrideparameters=no
timecolumn=TIME prefraction=0.5 postfraction=0.5  -V 5
epproc::hkgtigen:- hkgtigen 

Message of length 7519 truncated

Reply 1

Resend
From: Pedro M. Rodriguez Pascual <xmmhelp@sciops.esa.int>
To: nadine.may7@gmail.com
Subject: Re: EPICpn (PR#82016)
Date: Fri Jul 21 13:54:23 2017
Dear Nadine,

the warnings issued by epproc seems to suggest an inconsistency between the
calibration files (CCFs) and the SAS version.

Actually, you are using a rather old version of SAS (7.0.0 from 2006). Then we
have to encourage you move forward to the latest release SAS16
(https://www.cosmos.esa.int/web/xmm-newton/sas-download). Note also that a new
version SAS16.1 is expected to be released shortly (early August).

Another warning: only EPIC MOS cameras were operated during XMM-Newton
observation 0510011601 and they were set with calibration filters that block the
incident radiation from the telescopes. In brief, this observation does not
contain valid science data for its nominal pointing target XMMU J122939.7 . 

If you are interested on this target or source source with a few arcmin, please
try observation 0510011501 (or 0761630101, 0761630201 0761630301)

With best regards, 

Pedro

Up to top level
AO15   AO16   AO17   AO18   Backgrounds   Calibration   Conference   Data   Docs   EPICMOS   EPICpn   Feedback   Gallery   Misc   OM   Pending   PhD_Theses   Publications   RGS   RadMonitor   SAS_Hardware   SAS_WS   SASv16.0   SASv16.0_Installation   SASv16.1   SASv16.1_Installation   SASv17.0   SASv17.0_Installation   SASv18.0   SASv18.0_Installation   SciSim   Simulators_other   Suggestions   Trash   Visibility   XMM-bouncing   XMM-news   XRPS   XSA   esas   incoming  

Logged in as guest


Please make your (short) question the subject of your request!


Web interface using JitterBug ... back to the XMM home page